| General Data |
| Dimensions |
|
| Total Space |
54 x 63 cm |
| Height Without IV Pole |
133 cm |
| |
|
| Total Weigh |
approx. 73 kg |
| |
|
| Water Supply |
|
| Water Pressure |
min 20 psi, max 105 psi |
| Water Temperature |
min 10°C, max 25°C |
| Quality |
AAMI Standard, RD62:2006 |
| Quantity |
1 L/min |
| |
|
| Power Supply |
Singe phase AC 117 VAC |
| |
± 10%, 60 Hz, 12.5 A |
| Blood Circulation |
| Arterial Pressure0 |
|
| Prepump Indicating Range |
-300 to +280 mmHg |
| Postpump Indicating Range |
-80 to +500 mmHg |
| |
|
| Venous Pressure |
|
| Indicating Range |
-80 to +500 mmHg |
| |
|
| Transmembrane Pressure |
|
| Indicating Range |
60 to -520 mmHg |
| |
|
| Blood Pump |
|
| Delivery Range |
6 to 600 ml/min |
Internal Diameter of
Pump SegmenT |
2.6 4.8 6.4 8.0 mm |
| |
|
| SN Blood Pump |
|
| Delivery Range |
30 to 600 ml/min |
| |
|
| Air Detector |
Detection through ultrasonic transmission |
| |
|
| Heparin Pump
syringe |
10 to 20 ml disposable |
| Administration Rate |
0.1 to 9.9 ml/hr |
| Dialyzing Fluid |
| Quality |
AAMI Standard RD52:2004 |
| |
|
| Dialysate Flow |
100 to 800 ml/min |
| |
|
| Dialysate Temperature |
Selectable from 35°C to 39°C |
| |
|
| Dialysate Concentrate |
|
| Cond. Indicating Range |
10.0 to 17.0 mS/cm (25°C) |
| Alarm Limits Adjustable |
12.5 to 16.0 mS/cm (25°C) |
| |
|
| Ultrafiltration |
|
| UF-Rate (selectable) |
0 to 4.0 L/hr |
| UF-Factor of Dialyzers |
unlimited |
| |
|
| Blood Leak Detector |
0.45 ml/min (Hct 25%) |
|