TechnicalData

 

 
General Data
Dimensions  
Total Space 54 x 63 cm
Height Without IV Pole 133 cm
   
Total Weigh approx. 73 kg
   
Water Supply  
Water Pressure min 20 psi, max 105 psi
Water Temperature min 10°C, max 25°C
Quality AAMI Standard, RD62:2006
Quantity 1 L/min
   
Power Supply Singe phase AC 117 VAC
  ± 10%, 60 Hz, 12.5 A

 

Blood Circulation
Arterial Pressure0  
Prepump Indicating Range -300 to +280 mmHg
Postpump Indicating Range -80 to +500 mmHg
   
Venous Pressure  
Indicating Range -80 to +500 mmHg
   
Transmembrane Pressure  
Indicating Range 60 to -520 mmHg
   
Blood Pump  
Delivery Range 6 to 600 ml/min
Internal Diameter of
Pump SegmenT
2.6 4.8 6.4 8.0 mm
   
SN Blood Pump  
Delivery Range 30 to 600 ml/min
   
Air Detector Detection through ultrasonic transmission
   
Heparin Pump syringe 10 to 20 ml disposable
Administration Rate
0.1 to 9.9 ml/hr

 

Dialyzing Fluid
Quality AAMI Standard RD52:2004
   
Dialysate Flow
100 to 800 ml/min
   
Dialysate Temperature Selectable from 35°C to 39°C
   
Dialysate Concentrate  
Cond. Indicating Range 10.0 to 17.0 mS/cm (25°C)
Alarm Limits Adjustable 12.5 to 16.0 mS/cm (25°C)
   
Ultrafiltration  
UF-Rate (selectable) 0 to 4.0 L/hr
UF-Factor of Dialyzers unlimited
   
Blood Leak Detector 0.45 ml/min (Hct 25%)